Keywords: PIV flow measurement software; near-wall PIV measurement; Revealer RFlow software; wall-adaptive mesh deformation algorithm; fluid masking
The Revealer RFlow PIV flow field measurement software was used to process near-wall water flow particle images, examining the mesh layout and velocity field output of the wall-adaptive mesh deformation algorithm in the region near the wall, and comparing them with the results processed using a uniform grid. The image acquisition records were 1920×1080 px at 800 fps captured by a high-speed camera, comprising 8 groups with 2,400 frames each, processed using two-dimensional two-component (2D2C) PIV data processing. The quantitative assessment scope covered the near-wall band within 16 px of the wall, containing 12,286 evaluation points in total.
Relative to the same reference field, the root mean square error (RMSE) of the uniform grid and the RFlow software processing results were 3.243 and 0.590 px/frame, respectively — a relative reduction of approximately 81.8%. Wall-distance-binned statistics show that at smaller wall distances, the absolute deviation of the RFlow results exhibits a lower median and a narrower interquartile range.
The results indicate that the wall-adaptive mesh deformation enables the mesh morphology and measurement point locations to conform to the wall geometry; in this quantitative sample, the consistency of the RFlow results with the reference field is better than that of the uniform grid results.
Particle image velocimetry (PIV) obtains flow velocity information by estimating the displacement of tracer particle images between frames. Near a solid wall surface, the interrogation window may simultaneously cover both the fluid and solid regions; the solid side lacks effective particle signal, and wall reflection may also interfere with the cross-correlation computation. The combined effect of window truncation and velocity gradients may also cause the effective measurement location to deviate from the geometric center of the interrogation window.
For a uniform grid, the fixed measurement point layout is not necessarily suitable for the geometric conditions near the wall and at corners. After removing windows that straddle the solid body, measurement points may be missing in the wall-adjacent region. Even if the number of measurement points in the whole field increases, the near-wall region may still not obtain appropriate sampling locations.
The wall-adaptive mesh deformation algorithm adjusts the mesh morphology and measurement point locations according to the identified wall contour, so that the sampling layout adapts to local boundaries. Its direct effect is to improve the geometric fit between the mesh and the wall; it cannot be explained merely by an increase in the number of measurement points, nor can mesh deformation itself be equated to an improvement in effective spatial resolution.
This paper separately examines wall surface identification and mesh layout, near-wall velocity vector output, and the deviations of the two processing approaches relative to the same reference field on the same set of evaluation points.
The experimental material consists of near-wall particle images of water flow passing around a solid surface. A laser light sheet forms the measurement plane, and a high speed camera records the distribution of tracer particles within that plane; the software performs 2D2C PIV processing on the image pairs and outputs the two velocity components in the measurement plane.
Figure 1 is used to illustrate the layout relationship and does not serve as a basis for actual geometric dimensions or spatial calibration scale.

(Figure 1: Schematic diagram of the 2D2C PIV measurement layout — showing the experimental water channel, incident flow, bluff body (obstacle), laser light sheet (measurement plane), tracer particles, near-wall densified region, camera field of view (FOV), high-speed camera at 1920×1080 @ 800 fps, and the laser.)
Table 1 Image Data and Evaluation Conditions
Item | Record or Processing Condition |
Measurement type | Two-dimensional two-component PIV (2D2C) |
Image acquisition | 1920×1080 px; 800 fps |
Data volume | 8 groups, 2,400 frames per group |
Processing software | Revealer RFlow PIV flow field measurement software |
Comparison scheme | Uniform grid vs. wall-adaptive mesh deformation algorithm |
Evaluation scope | d ≤ 16 px; evaluation points in the same batch, N = 12,286 |
Statistical comparison unit | px/frame |
Reference field | Case U = 0.3 m/s, Re = 3000 |
The curved-surface near-wall evaluation points, velocity scatter plots, and wall-distance statistical plots serve as the primary quantitative cases; the RMSE and the 81.8% relative reduction are attributed solely to that set of evaluation points. Another set of boundary contours, mesh plots, and vector plots is used to illustrate how the RFlow software determines the fluid domain and arranges near-wall sampling points, and does not serve as a second set of quantitative validation.
Table 2 Purposes and Metric Attribution of the Two Sets of Demonstration Materials
Material Category | Figures | Report Purpose |
Primary quantitative case | Figure 4 evaluation points and velocity scatter; Figure 5 wall-distance statistics | Comparison of the two processing results on the same batch of evaluation points |
Method demonstration case | Figure 2 boundary and mesh; Figure 3 velocity vectors | Illustration of boundary, sampling locations, and vector coverage |
The two sets of materials cannot be pieced together as a before-and-after processing of the same case. The 8 groups of data in the acquisition records and the 12,286 evaluation points also do not belong to the same statistical concept: the former describes the acquisition scale, and the latter describes the evaluation set.
The RFlow software first identifies the solid boundary and generates a fluid-domain mask to determine the effective fluid domain for velocity computation. Subsequently, the wall-adaptive mesh deformation algorithm adjusts the mesh morphology and measurement point locations according to the local wall contour, so that the sampling layout near the wall surface and corners conforms to the geometric boundary. Boundary identification provides the geometric constraint of the fluid domain, while mesh deformation adjusts the sampling locations.
Figure 2 shows the wall contour in the particle images and the corresponding mesh distribution. Near the boundary, it can be observed that the mesh adjusts along the wall contour, providing a basis for understanding the spatial positions of adjacent measurement points.
(Figure 2: Boundary identification and mesh distribution in the method demonstration materials — (a) boundary contour and particle image; (b) wall-following mesh distribution. The red line in the left image marks the wall contour; the right image shows the differences in sampling spacing between the near-wall and far-field regions.)
The quantitative comparison uses the same batch of evaluation points within the same near-wall band, with both processing results compared against the same reference field. The wall distance d is expressed in pixels, and the velocity statistics are retained in px/frame. RMSE is used to describe the root mean square deviation between the selected velocity evaluation quantity and the reference values.
Figure 3 shows the vector locations below the wall surface and near the corners, providing sampling information for observing local velocity variations. This figure is used to demonstrate the output of the RFlow software and is not paired with an independent uniform-grid vector plot for co-located comparison; the quantitative comparison of the two methods is presented in the subsequent statistics.

(Figure 3: Near-wall velocity vector display from the RFlow software, intended for observing sampling locations and vector distribution.)
The relationship between the evaluation point locations within the curved-surface near-wall band and the velocity scatter is shown in Figure 4. The RFlow point cloud is overall closer to the 1:1 reference line, while the uniform-grid point cloud has a larger dispersion range. The 1:1 line indicates that the output velocity magnitude equals the reference velocity magnitude, and is used to observe the consistency between the two.
(Figure 4: Near-wall evaluation points and velocity scatter corresponding to the quantitative sample — (a) near-wall evaluation point locations (color = wall distance d, d ≤ 16 px, N = 699/4); (b) same batch of near-wall evaluation points: output vs. ground truth. Conventional PIV (uniform grid): RMSE = 3.243, R² = 0.192; RFlow adaptive PIV (wall-adapted densification): RMSE = 0.590, R² = 0.91. “Ground truth” in the figure refers to the reference values.)
Table 3 RMSE Comparison Relative to the Reference Field
Metric | Uniform Grid | RFlow Wall-Adaptive Mesh Deformation |
RMSE (px/frame) | 3.243 | 0.590 |
RMSE relative reduction | Used as the comparison baseline | Approximately 81.8% |
This reduction indicates that, under the conditions of this evaluation point set and the reference field, the RFlow results have a lower root mean square deviation.
Figure 5 shows, grouped by wall distance, the median and the 25%–75% interquartile range of the absolute deviation. At smaller wall distances, the RFlow curve lies below the uniform-grid curve, and the shaded band is also narrower; as the wall distance increases, the deviations of both methods generally decrease. This indicates that in this case, the differences are mainly concentrated in the wall-adjacent region.

(Figure 5: Variation of the absolute deviation relative to the reference field with wall distance. Solid lines are medians; shaded areas are the 25%–75% interquartile ranges.)
The method demonstration results indicate that the wall-adaptive mesh deformation algorithm of the Revealer RFlow software can adjust the mesh morphology and measurement point locations according to the wall contour, providing sampling information for observing the velocity distribution near the wall surface and corners.
In another curved-surface quantitative case, on the 12,286 near-wall evaluation points, the RMSE decreased from 3.243 to 0.590 px/frame, a relative reduction of approximately 81.8%; wall-distance binning shows that at smaller wall distances, the median of the absolute deviation is lower and the interquartile range is narrower.
The wall-adaptive mesh deformation algorithm of the Revealer RFlow software effectively improves the sampling fit between the fixed grid and the wall geometry, making the measurement point layout near the wall surface and corners better conform to the fluid boundary, and provides a local measurement basis for near-wall velocity distribution analysis in bluff-body flows, boundary layers, and separated flows.
English